Titrant | Specifics | Electrode description* | Cat. No./ Order No. | Reference Electrolyte | Remarks |
Aqueous acid/bases | | Combined LL pH glass electrode | 020912146 (6.0232.100) | c(KCl) = 3 mol/L 020096101 (6.2308.020) | |
alkaline samples | Combined LL pH glass electrode, U glass | 020912162 (6.0222.100) | c(KCl) = 3 mol/L 020096101 (6.2308.020) | |
titrations performed at high temperatures | Unitrode | 020910674 (6.0259.100) | c(KCl) = 3 mol/L 020096101 (6.2308.020) | |
titrations with small sample volumes | Combined LL micro pH glass electrode OR Combined flat membrane pH glass electrode | 020912235 (6.0234.100) 020915617 (6.0227.100) | c(KCl) = 3 mol/L 020096101 (6.2308.020) | Shaft diameter 6.4 mm, minimal immersion depth; 10 mm Shaft diameter 12 mm, minimal immersion depth; 1–2 mm |
titrations in water with low conductivity | Aquatrode plus | 020910631 (6.0252.300) | c(KCl) = 3 mol/L 020096101 (6.2308.020) | Typical application: water analysis - p&m value etc. |
Perchloric acid | See AB 80 and Monograph"Non-aqueous titration of acids and bases with potentiometric endpoint indication" | Solvotrode - combined LL pH glass electrode, ground-joint diaphragm (PCTFE) | 020910551 (6.0229.100) | c(LiCl) = 2 mol/L in Ethanol 020094435 (6.2312.010) | easy to clean sleeve diaphragm IMPORTANT: When not in daily use, store in LiCI sat. in ethanol and soak membrane in distilled water before use. However, when in daily use, store in distilled water, so that only the membrane is immersed, not the diaphragm. Before the measurement lift the diaphragm sleeve slightly and allow a small amount of electrolyte to flow out. Rinse the electrode with distilled water, replenish the reference electrolyte and start the titration (see bottom of chart). IMPORTANT: Never fill the Solvotrode with LiCIO4 in glacial acetic acid! |
TBAOH | See AB 80 and Monograph"Non-aqueous titration of acids and bases with potentiometric endpoint indication" | Solvotrode - combined LL pH glass electrode, ground-joint diaphragm (PCTFE) | 020910551 (6.0229.100) | c(TEABr) = 0.4 mol/L in ethylen glycol1 020094451 (6.2320.000) |
Potassium methylate | | Solvotrode - combined LL pH glass electrode, ground-joint diaphragm (PCTFE) | 020910551 (6.0229.100) | c(TEABr) = 0.4 mol/L in ethylen glycol1 020094451 (6.2320.000) |
Cyclohexyl-amine | | Solvotrode - combined LL pH glass electrode, ground-joint diaphragm (PCTFE) | 020910551 (6.0229.100) | saturated LiCl in Ethanol 020094443 (6.2312.000) |
Silver nitrate | AB 25 | Ag-Titrode, possibly Ag2S coated | 6.0430.100 | | Electrode is also available with a plug-in head S: 20 94 852-3 (6.0430.500)3 |
arsenite, cerium sulphate, iron (III), iodine, potassium bromate, sodium nitrite, sodium thiosulphate, oxalic acid, permanganate, titanium | For titrations where the pH value during the titration doesn't change. | Pt-Titrode | 6.0431.100 | | Electrode is also available with a plug-in head S: 20 94 852-3 (6.0431.500)3 |
For titrations during which the pH value changes. | Combined LL Pt-ring electrode | 6.0451.100 | c(KCl) = 3 mol/L 020096101 (6.2308.020) | |
Karl-Fisher reagent | | Double Pt-wire electrode | 020928841 (6.0338.100) | | |
Cr3+ , Fe2+, | Also for COD titrations | Micro Au-Titrode | 020948604 (6.0435.110) | | L = 160 mm, Loger version available: (6.9904.023), L= 250 mm |
EDTA | Back-titration of the BaCl2 excess. For sulphate determination see AB 140. | Polymer membrane electrode, Ca2+ | 020955856 (6.0508.110) | c(KCl) = 3 mol/L 020096101 (6.2308.020) | a separate reference electrode is needed. Order no: 20 94 411-1 (6.0726.107) |
EDTA Complexon® III and IV | Determination of Al, Ba, Ca, Cd, Bi, Co, Fe, Mg, Ni, Pb, Zn. See AB 125 | Ion selective electrode Cu2+ | 020959002 (6.0502.140) | c(KCl) = 3 mol/L 020096101 (6.2308.020) | a separate reference electrode is needed. Order no: 20 94 411-1 (6.0726.107) |
Determination of Ca, Mg, See AB 125 | Polymer membrane electrode, Ca2+ | 020955856 (6.0508.110) | | |
Surfactants in aqueous media | Tego® trant A 100,2 see AB 233 | Ionic surfactant electrode | 020959321 (6.0507.120) | c(KCl) = 3 mol/L 020096101 (6.2308.020) | a separate reference electrode is needed. Order Nr: 6.0726.100 |
sodium lauryl sulphate, see AB 233 | Ionic surfactant electrode | 020959321 (6.0507.120) | c(KCl) = 3 mol/L 020096101 (6.2308.020) | a separate reference electrode is needed. Order Nr: 6.0726.100 |
Sodium tetra phenyl borate, see AB 230, AB 263 | NIO surfactant electrode | 020959304 (6.0507.010) | Inner chamber of the reference electrode: c(KCI) = 3 mol/L 020096101 (6.2308.020) Outer chamber: c(NaCL) = 1 mol/L | a separate reference electrode is needed. Order Nr: 6.0726.107 |
Non aqueous surfactants | Nonaqueous media, solvents as Chloroform, MIKB, see AB 269 | Surfactrode resistant | 020959312 (6.0507.130) | c(KCl) = 3 mol/L 020096101 (6.2308.020) | a separate reference electrode is needed. Order no: 20 94 411-1 (6.0726.107) electrode not suitable for high salt concentrations and pH > 10 |
Nonaqueous media, solvents as MIKB Two phase titration, see AB 269 | Surfactrode refill | 020959339 (6.0507.140) | c(KCl) = 3 mol/L 020096101 (6.2308.020) | a separate reference electrode is needed. Order no: 20 94 411-1 (6.0726.107) electrode not suitable for Chloroform |