Determination of Aluminium with Potentiometric Method
Equipment
Potentiometer = 751 GPD Titrino / Titroprocessor
Electrodes = 6.0232.100 combined pH glass electrode; input 1
Reagents
c(HCl) = 1 mol/L; D0
c(NaOH) = 1 mol/L; D1
w(KF) = 0.10 (10%); D2
Sample
a sample with 40...50 mg Al content
50 mL dist. water
5 mL w(sodium gluconate) = 0.25 (25%)
Remarks
Determination reaction:
Al(OH)3 + 6 KF K3AlF6 + 3 OH-
Calculations:
Aluminum=(C70+C71)*C01*C02*C03/C04/C00;3;%
Aluminum = content of Al in %
C01 = concentration of titrating agent (1 mol/L)
C02 = molecular mass / “normality“ (26.982 / 3 = 8.994)
C03 = factor for % (100)
C04 = factor for mL L (1000)
C70 = excess of the method “Al 2“ (stop V - EP1)
C71 = consumption in method “Al 4“ (EP1)
Answer
Literature
Metrohm Application Bulletin No. 24: Potentiometric determination of aluminium
Method documentation
'pa
751 GPD Titrino 15215 751.0010
TIP Aluminum
parameters
>sequence
1.stirrer: …ON
2.method: …Al 1
3.method: …Al 2
4.method:… Al 3
5.pause …180 s
6.method: …Al 4
>statistics
status: …OFF
>preselections
req.ident:… OFF
req.smpl size:… value
meas.mode: …pH
meas.input: ….1
temperature …25.0 °C
------------
'fm
751 GPD Titrino 15215 751.0010
TIP Aluminum
>calculations
Aluminum=(C70+C71)*C01*C02*C03/C04/C00;3;%
C00= 0.2986
C01= 1
C02= 8.994
C03= 100
C04= 1000
C70= 0.193
C71= 2.684
------------
'fr
751 GPD Titrino 15215 751.0010
TIP Aluminum
smpl size …0.2986 g
C70 ….0.193
C71 …..2.684
Aluminum….. 8.666 %
TIP terminated
============